UNICEL DXC 600 PRO SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2014-00501
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 22, 2014
- Report Date
- October 22, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- JGS
- PMA / PMN Number
- K103842
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER CONFIRMED CONTAMINATION IN THE FLOW CELL BASED ON THE CULTURE RESULT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND EVALUATED THE DEVICE. THE FSE PERFORMED PM C AND PM D [PREVENTIVE MAINTENANCE, ION-SELECTIVE ELECTRODE (ISE) AND MODULAR CHEMISTRY (MC) CUPS] AND REPLACED MULTIPLE ISE HARDWARE COMPONENTS [RATIO PUMP, REFERENCE ELECTRODE, ISE TUBING, EIC (ELECTROLYTE INJECTION CUP) VALVES, EIC DIAPHRAGM AND CARBON BRIDGE] TO RESOLVE THE ISSUE.
THE CUSTOMER REPORTED THAT THE UNICEL DXC 600 PRO SYNCHRON SYSTEM GENERATED TWO FALSE HIGH SODIUM (NA) RESULTS. THE CUSTOMER VERBALLY PROVIDED SODIUM RESULTS FROM TWO PATIENTS. THE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO IMPACT TO PATIENT TREATMENT. THE CUSTOMER WAS NOT AWARE OF ANY PATIENT TREATMENT AFFECTED. CONTROLS (QC) RUN BEFORE AND AFTER THE EVENT WERE WITHIN LAB-ESTABLISHED RANGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 738654 | UNICEL DXC 600 PRO SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JGS | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |