FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4252180 · Received November 14, 2014

Report

Report Number
2032227-2014-52922
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 18, 2014
Report Date
October 18, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AND IT ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED AND OR LOOSE DRIVE SUPPORT DISK. COMPROMISED FORCE SENSOR SYSTEM ALARM WAS NOT VERIFIED DURING TESTING DUE TO LOOSE DRIVE SUPPORT DISK. THE DEVICE WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CASE CRACKED AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NUMBERS ON THE INSULIN PUMP DO NOT GO UP. CUSTOMER STATED THAT THEY DELIVERED 30 UNITS OF INSULIN WITHOUT NUMBERS SHOWING UP ON THE SCREEN. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 204MG/DL. TROUBLESHOOTING OCCURED. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737182 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 21 YR