PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-52914
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 18, 2014
- Report Date
- October 18, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
NO BUTTON ERROR ALARM WAS NOTED. THE ACT AND ESCAPE BUTTONS DID NOT RESPOND DUE TO CORRODED KEYPAD TRACES. NO DISPLAY ANOMALY WAS NOTED. NO ERROR ALARM COULD BE VERIFIED DUE TO THE UNRESPONSIVE BUTTONS. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, AND A CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM AS WELL AS AN UNEXPECTED RESTART ALARM. IT WAS ALSO REPORTED THAT THE NUMBERS ON THE INSULIN PUMP SCROLL UP WITHOUT ANY INPUT FROM THE CUSTOMER. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 372MG/DL. TROUBLESHOOTING OCCURRED. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 738418 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |