PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-52908
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 18, 2014
- Report Date
- October 18, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)
NO BATTERY OUT LIMIT NOTED. INSULIN PUMP RECEIVED WITH NORMAL OPERATING CURRENTS. THE BATTERY ALARM FUNCTIONING PROPERLY. INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED UP AND DOWN BUTTON ON KEYPAD TRACES. INSULIN PUMP HAD BROKEN BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT THE NUMBERS ON THE INSULIN PUMP SCROLL UP WITHOUT ANY INPUT FROM THE CUSTOMER. IT WAS ALSO REPORTED THAT THE CUSTOMER RECEIVED A BATTERY OUT LIMIT ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 155MG/DL. TROUBLESHOOTING OCCURRED. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736816 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |