ANIMAS VIBE
Report
- Report Number
- 2531779-2014-32708
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Report Date
- October 28, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2015-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING A VISUAL INSPECTION OF THE PUMP, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. IT WAS REPORTED THAT THE PUMP BATTERY COMPARTMENT WAS DAMAGED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 738169 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |