FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4251737 · Received November 14, 2014

Report

Report Number
2032227-2014-30147
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
May 22, 2014
Report Date
October 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARMS NOTED. MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED FOR MOTOR ERROR DURING THE BOLUS PROCESS. THE REPORTER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ALARM. ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
735992 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-554LWWB