FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT EZ

MDR report key: 4251656 · Received November 14, 2014

Report

Report Number
1826988-2014-00410
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 17, 2014
Report Date
October 24, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K111268
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CONTOUR NEXT TEST STRIP INFORMATION WAS NOT PROVIDED IN THE INITIAL REPORT. THE TEST STRIPS WERE RETURNED FOR EVALUATION. LOT# 4EFEB07, EXPIRATION DATE 05/31/2016. MANUFACTURE DATE 05/01/2014.

Description of Event or Problem · 1

THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF LO ON THE CONTOUR NEXT EZ, RE-TESTED ON A DIFFERENT METER AND THE READING WAS 133MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE TEST STRIP INFORMATION WAS INVALID. A NEW METER WAS SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736411 CONTOUR NEXT EZ BLOOD GLUCOSE METER NBW BAYER HEALTHCARE LLC 7252

Patients

Seq Age Sex Outcome Treatment
1