FDA Adverse Event
Malfunction
Summary report: N
CONTOUR NEXT EZ
MDR report key: 4251656
·
Received November 14, 2014
Report
- Report Number
- 1826988-2014-00410
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 17, 2014
- Report Date
- October 24, 2014
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K111268
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CONTOUR NEXT TEST STRIP INFORMATION WAS NOT PROVIDED IN THE INITIAL REPORT. THE TEST STRIPS WERE RETURNED FOR EVALUATION. LOT# 4EFEB07, EXPIRATION DATE 05/31/2016. MANUFACTURE DATE 05/01/2014.
Description of Event or Problem · 1
THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF LO ON THE CONTOUR NEXT EZ, RE-TESTED ON A DIFFERENT METER AND THE READING WAS 133MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE TEST STRIP INFORMATION WAS INVALID. A NEW METER WAS SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736411 | CONTOUR NEXT EZ | BLOOD GLUCOSE METER | NBW | BAYER HEALTHCARE LLC | 7252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |