FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 4251655 · Received November 14, 2014

Report

Report Number
1045834-2014-14782
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 25, 2014
Report Date
October 27, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE MOTOR DOES NOT RUN DUE TO A WORN THERMISTOR, WHICH GENERATED AND E6 ERROR FROM REPEATED OVERHEATING OF THE DEVICE. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE, ABUSE AND/OR ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CRANIOTOMY SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE DISPLAYED AN E6 ERROR CODE AND WAS OVERHEATING. AS A RESULT, THERE WAS A FIVE-MINUTE DELAY TO THE SURGICAL PROCEDURE TO SET UP AN IDENTICAL SPARE DEVICE. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738544 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1