FDA Adverse Event Malfunction Summary report: N

TERUMO SURGUARD2 SAFETY SYRINGE WITH NEEDLE

MDR report key: 4251616 · Received November 14, 2014

Report

Report Number
3003902955-2014-00024
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 16, 2014
Report Date
November 14, 2014
Manufacturer
TERUMO PHILIPPINES CORPORATION
Product Code
MEG
PMA / PMN Number
K051865
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETENTION AND CURRENTLY LOT SAMPLES WERE EVALUATED. VISUAL INSPECTION CONFIRMED THE FOLLOWING: (1) THE COLLAR, SAFETY SHEATH AND THE SHEATH TOOTH OF ACTUAL AND RETENTION SAMPLES WERE FREE FROM MOLDING RELATED DEFECTS: (2) THE SAFETY SHEATH WAS PROPERLY ATTACHED ON THE ASSEMBLE NEEDLE AND THE COLLAR WAS FULLY SEATED ON THE HUB; (3) THE SAFETY SHEATH WAS PROPERLY ATTACHED ON THE ASSEMBLED NEEDLE AND THE COLOR WAS FULLY SEATED ON THE HUB, WHICH INDICATES THE SHEATH WAS IN NORMAL ALIGNMENT. FUNCTION TESTING CONFIRMED THAT PERFORMANCE SPECIFICATIONS WERE MET. SIMULATION TESTING WAS CONDUCTED. ACTUAL SAMPLES WERE MANUALLY ACTIVATED AS PER INSTRUCTIONS FOR USE (IFU), BY (C) FINGER ACTIVATION, (D) THUMB ACTIVATION AND (E) SURFACE ACTIVATION. AUDIBLE SOUND (CLICK) WAS HEARD AND THE NEEDLE WAS COMPLETELY ENGAGED UNDER THE TOOTH UPON VISUAL INSPECTION. THE NEEDLE WAS COMPLETELY ENGAGED UNDER THE TOOTH. SUCCESSFUL ACTIVATION WAS ACHIEVED WITHOUT BREAKING THE SAFETY SHEATH AND NO DIFFICULTY WAS EXPERIENCED. A REVIEW OF THE DEVICE HISTORY INSPECTION RECORDS INDICATED THAT THERE WERE NO PRODUCTION RELATED PROBLEMS. ALTHOUGH THE CAUSE OF THE REPORTED EVENT CANNOT BE DEFINITIVELY DETERMINED BASED UPON THE AVAILABLE INFORMATION, THERE IS NO EVIDENCE THAT THIS EVENT WAS RELATED TO A DEVICE DEFECT OR MALFUNCTION. THE DEVICE LABELING DOES ADDRESS THE POTENTIAL FOR SUCH AN OCCURRENCE IN THE WARNINGS/CAUTIONS AND THE INSTRUCTIONS-FOR-USE WITH STATEMENTS SUCH AS THE FOLLOWING: (1) "POSITION SHEATH APPROXIMATELY 45 DEGREES TO A FLAT SURFACE. PRESS DOWN WITH A FIRM, QUICK MOTION UNTIL A DISTINCT AUDIBLE CLICK IS HEARD"; AND (2) "VISUALLY CONFIRM THAT THE NEEDLE IS FULLY ENGAGED UNDER THE LOCK". ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY ASSURANCE FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW-UP.

Additional Manufacturer Narrative · 1

(B)(4). THE INVOLVED DEVICE WAS NOT RETURNED FOR EVALUATION, WHICH LIMITS THE FAILURE INVESTIGATION TO ASSESSMENT OF USER FACILITY INFORMATION AND EVALUATION OF RESERVE SAMPLES FROM THE REPORTED LOT AND CURRENT LOT SAMPLES. NO ABNORMALITIES OR DEFECTS WERE NOTED ON VISUAL INSPECTION AND ALL SAMPLES PASSED FUNCTIONAL TESTING. SIMULATION TESTING WAS PERFORMED, RETENTION SAMPLES WERE MANUALLY ACTIVATED AS PER INSTRUCTIONS FOR USE (IFU), BY PRESSING DOWN WITH A FIRM AND QUICK MOTION ON A FLAT SURFACE (SHEATH IS POSITIONED APPROXIMATELY 450). AUDIBLE SOUND (CLICK) WAS HEARD AND THE NEEDLE WAS COMPLETELY ENGAGE UNDER THE TOOTH UPON VISUAL CONFIRMATION. SUCCESSFUL ACTIVATION WAS ACHIEVED WITHOUT ANY ABNORMALITIES. A REVIEW OF THE DEVICE HISTORY RECORDS AND INSPECTION RECORDS INDICATED THAT THERE WERE NO PRODUCTION RELATED PROBLEMS. ALTHOUGH THE CAUSE OF THE REPORTED EVENT CANNOT BE DEFINITIVELY DETERMINED BASED UPON THE AVAILABLE INFORMATION, THERE IS NO EVIDENCE THAT THIS EVENT WAS RELATED TO A DEVICE DEFECT OR MALFUNCTION. THE DEVICE LABELING DOES ADDRESS THE POTENTIAL FOR SUCH AN OCCURRENCE IN THE WARNINGS/CAUTIONS AND THE INSTRUCTIONS-FOR-USE WITH STATEMENTS SUCH AS THE FOLLOWING: (1) "HANDLE WITH CARE TO AVOID NEEDLESTICKS"; (2) "KEEP HANDS BEHIND THE NEEDLE AT ALL TIMES DURING USE AND DISPOSAL"; (3) "FOR GREATEST SAFETY, ACTIVATE THE SHEATH USING A ONE-HANDED TECHNIQUE"; (4) "IF A NEEDLE IS BENT OR DAMAGED, NO ATTEMPT SHOULD BE MADE TO STRAIGHTEN NEEDLE OR USE THE PRODUCT"; (5) "POSITION SHEATH APPROXIMATELY 45 DEGREES TO A FLAT SURFACE. PRESS DOWN WITH A FIRM, QUICK MOTION UNTIL A DISTINCT AUDIBLE CLICK IS HEARD"; (6) "VISUALLY CONFIRM THAT THE NEEDLE IS FULLY ENGAGED UNDER THE LOCK"; AND (7) "DISPOSE OF USED NEEDLES AND MATERIALS FOLLOWING THE POLICIES AND PROCEDURES OF YOUR FACILITY AS WELL AS FEDERAL AND LOCAL REGULATIONS FOR SHARPS DISPOSAL." ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY ASSURANCE FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW-UP. (B)(4). DEVICE NOT RETURNED.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT A NURSE INCURRED A NEEDLE STICK AFTER AN INJECTION. FOLLOW-UP COMMUNICATION WITH THE USER FACILITY CONFIRMED; (1) AN INJECTION OF ANTIBIOTIC WAS GIVEN TO THE PATIENT; (2) AFTER THE INJECTION THE NURSE PUT ON THE SHEATH AND LAID THE DEVICE DOWN; (3) WHEN THE NURSE PICKED IT UP THE SHEATH FELL OFF; (4) THE NEEDLE STUCK THE NURSE ON THE PALM OF HER RIGHT HAND BETWEEN THE THIRD AND FOURTH FINGERS; AND (5) DIAGNOSTIC BLOOD TESTING IS BEING PERFORMED ON BOTH THE NURSE AND PATIENT.

Description of Event or Problem · 1

THIS REPORT IS BEING SUBMITTED AS FOLLOW-UP #1 FOR MFG. REPORT # 3003902955-2014-00024 TO PROVIDE RESULTS OF THE RETENTION AND CURRENT LOT SAMPLE EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736344 TERUMO SURGUARD2 SAFETY SYRINGE WITH NEEDLE TERUMO NEOLUS NEEDLE MEG TERUMO PHILIPPINES CORPORATION NA 140602A

Patients

Seq Age Sex Outcome Treatment
1