FDA Adverse Event
Malfunction
Summary report: N
TRUERESULT
MDR report key: 4251117
·
Received November 5, 2014
Report
- Report Number
- 1052693-2014-00489
- Event Type
- Malfunction
- Date Received
- November 5, 2014
- Date of Event
- October 7, 2014
- Report Date
- January 4, 2016
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND ON METER. DEFECT FOUND ON STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: STRIPS HAD IMPROPER STORAGE.
Additional Manufacturer Narrative · 1
PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).
Description of Event or Problem · 1
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO". LAST 5 TESTS IN MEMORY: "LO", "LO", "LO", 140MG/DL, 101MG/DL. CUSTOMER NORMALLY READS BETWEEN 115-140MG/DL. NO ADVERSE EVENT REPORTED.
Description of Event or Problem · 1
CONSUMER COMPLAINT OF BLOOD RESULT READING OF "LO". LAST 5 TESTS IN MEMORY: "LO", "LO", "LO", 140MG/DL, 101MG/DL. CUSTOMER NORMALLY READS BETWEEN 115-140MG/DL. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 710444 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1622 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |