FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4251051 · Received November 14, 2014

Report

Report Number
2032227-2014-25835
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 21, 2014
Report Date
October 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: UNIT PASSED REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33,EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TEST. NO MOTOR ERROR ALARM. MOTOR PASSED MOTOR TEST. MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING PRIME. CUSTOMER'S BLOOD GLUCOSE LEVELS WERE NOT INCLUDED IN THE REPORT. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. IT WAS REPORTED THAT CUSTOMER WAS ABLE TO REWIND THE INSULIN PUMP. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738526 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 22 YR