SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-40573
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 20, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 ALARM (AIR IN SET/LINE) OCCURRED ON THE HOMECHOICE DEVICE DURING THE INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED TO THE PATIENT THAT AIR HAS ENTERED THE SETUP AND INSTRUCTED THE PATIENT TO CYCLE THE POWER TO CLEAR THE ALARM. THE TSR ADVISED THE PATIENT TO START OVER WITH ALL NEW SUPPLIES. THERE WAS NOTHING UNUSUAL FOUND DURING THE TROUBLESHOOTING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736291 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | HOMECHOICE |