FDA Adverse Event Malfunction Summary report: N

VITROS 5,1 FS CHEMISTRY SYSTEM

MDR report key: 4250697 · Received November 14, 2014

Report

Report Number
1319681-2014-00183
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 9, 2014
Report Date
November 14, 2014
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT LOWER AND HIGHER THAN EXPECTED VITROS AMON RESULTS WERE OBTAINED FROM TWO QUALITY CONTROL FLUIDS AND A PROFICIENCY FLUID PROCESSED ON A VITROS 5,1 FS CHEMISTRY SYSTEM. THE MOST LIKELY ASSIGNABLE CAUSE IS INSTRUMENT RELATED, AS A WITHIN-RUN PRECISION TEST WAS OUTSIDE OF OCD GUIDELINES, INDICATING THE VITROS 5,1 FS CHEMISTRY SYSTEM WAS NOT PERFORMING AS INTENDED. FOLLOWING SERVICE ACTIONS, PERFORMANCE WAS VERIFIED BY PERFORMING PRECISION TESTING THAT WAS WITHIN THE ACCEPTABLE GUIDELINES INDICATING THE VITROS 5,1 FS CHEMISTRY SYSTEM WAS PERFORMING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT LOWER AND HIGHER THAN EXPECTED VITROS AMON RESULTS WERE OBTAINED FROM TWO QUALITY CONTROL FLUIDS AND A PROFICIENCY FLUID PROCESSED ON A VITROS 5,1 FS CHEMISTRY SYSTEM. PROFICIENCY SAMPLE - 125 VS. EXPECTED 167.8 UMOL/L; QC FLUID 1 - 97.4 VS. EXPECTED 74.5 UMOL/L; QC FLUID 2 - 164.1, 135.9, 163.0 VS. EXPECTED 225.0 UMOL/L. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION, SHOULD THEY OCCUR UNDETECTED ON PATIENT SAMPLES. THE CUSTOMER HAS CONFIRMED THAT NO REPORTED PATIENT AMON RESULTS HAD BEEN IN QUESTION. HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WERE NOT AFFECTED OR WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736639 VITROS 5,1 FS CHEMISTRY SYSTEM CHEMISTRY ANALYZER JJE ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1