ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-32655
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Report Date
- October 28, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 01/29/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/10/2015 WITH THE FOLLOWING FINDINGS: NO DEFECT WAS FOUND. UNABLE TO DUPLICATE THE COMPLAINT.THE BLACK BOX SHOWS AN UNCONFIRMED REPLACE CARTRIDGE ON 10/28/2014; ALARM WAS UNCONFIRMED FOR 5HRS 49MINS. THE UNCONFIRMED ALARM BYPASSED THE LOW BATTERY WARNING RESULTING IN AN EAW-128 REPLACE BATTERY ALARM AT 06:36. PUMP WAS EXERCISED FOR 24HRS WITH NO EAW¿S OCCURRING. BATTERY CAP AND CARTRIDGE CAP WERE NOT RETURNED; TEST CAPS USED TO COMPLETE TESTING. A LOW BATTERY WARNING WAS REPRODUCED ON THE BENCH FOR TESTING PURPOSES; THE WARNING WAS RECORDED AT THE CORRECT VOLTAGE THRESHOLD AND WAS ACCURATELY RECORDED IN THE PUMP HISTORY.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (HISTORY/SETTINGS ISSUE) ISSUE. REPORTER WAS CONCERNED THAT NO LOW BATTERY WARNING APPEARED IN PUMP HISTORY AND PUMP DID NOT ALARM THAT THERE IS A LOW BATTERY BEFORE PUMP SHUT OFF. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 738685 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |