FDA Adverse Event Malfunction Summary report: N

SMALL BATTERY DRIVE 14.4V BATTERY

MDR report key: 4250558 · Received November 14, 2014

Report

Report Number
2520274-2014-14746
Event Type
Malfunction
Date Received
November 14, 2014
Report Date
October 29, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
MOQ
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE WOULD HOLD A CHARGE. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOCATION WAS UNKNOWN. DEVICE MANUFACTURE DATE IS UNKNOWN. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE WAS LOSING POWER. THE EVENT WAS NOT RELATED TO SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. SEVERAL ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED IF FURTHER INFORMATION IS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738666 SMALL BATTERY DRIVE 14.4V BATTERY BATTERY, REPLACEMENT, RECHARGEABLE MOQ DEPUY SYNTHES POWER TOOLS D130411

Patients

Seq Age Sex Outcome Treatment
1