FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4250497 · Received November 14, 2014

Report

Report Number
1416980-2014-40348
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 17, 2014
Report Date
October 17, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. THE F-38 ALARM WAS IDENTIFIED DURING THE FUNCTIONAL TESTING. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE DAMAGED FORCE SENSING RESISTORS (FSRS). TO CORRECT THE CONDITION, THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO PRESENT AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736415 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1