SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-40453
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 20, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN LINE/SET) OCCURRED DURING DWELL THREE OF THREE ON THE HOMECHOICE. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THERE WAS NOTHING FOUND DURING TROUBLESHOOTING THAT WOULD CAUSE OR CONTRIBUTE TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE HP TO CLEAR THE ALARM AND END THERAPY. THE HP INDICATED THEY WOULD COMPLETE THERAPY WITH MANUAL SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736598 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | HOMECHOICE |