FDA Adverse Event Malfunction Summary report: N

ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT

MDR report key: 4250264 · Received November 6, 2014

Report

Report Number
1419322-2014-00073
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 7, 2014
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
EKX
PMA / PMN Number
K031145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. BUR END BEARING RETAINER FAILURE, FREE ROAMING BALL BEARINGS AND EXCESSIVE DEBRIS MOST LIKELY CREATED FRICTION WITHIN THE HEAD WHICH RESULTED IN TEMPERATURE INCREASE. IT IS REASONABLE TO SUSPECT GEAR FUNCTION WAS COMPROMISED BY THE ADDED DEBRIS INSIDE THE HEAD WHICH IS EVIDENT FROM ABRASIVE WEAR ON THE TEETH OF THE GEARS. ROOT CAUSE CODE: SET/TAIL/HEAD ASSEMBLY, EXCESSIVE USE/ABUSE.

Description of Event or Problem · 1

IN THIS EVENT A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714285 ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT HANDPIECE, DIRECT DRIVE, AC-POWERED EKX DENTSPLY PROFESSIONAL NA