FLOGARD
Report
- Report Number
- 1416980-2014-40260
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 21, 2014
- Report Date
- October 21, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
COMPLAINT NO: CMPLNT-(B)(4). (B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, POWER ON SELF TESTING, BATTERY TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. VISUAL INSPECTION NOTED A DAMAGED BATTERY. A REVIEW OF THE ALARM LOG AND POWER ON SELF TESTING IDENTIFIED AN F-94 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A DAMAGED BATTERY. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLO-GARD INFUSION PUMP PRESENTED AN F-94 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.