FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4250255 · Received November 14, 2014

Report

Report Number
1416980-2014-40260
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 21, 2014
Report Date
October 21, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). (B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, POWER ON SELF TESTING, BATTERY TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. VISUAL INSPECTION NOTED A DAMAGED BATTERY. A REVIEW OF THE ALARM LOG AND POWER ON SELF TESTING IDENTIFIED AN F-94 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A DAMAGED BATTERY. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLO-GARD INFUSION PUMP PRESENTED AN F-94 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.