FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4250250 · Received November 14, 2014

Report

Report Number
1416980-2014-40246
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 18, 2014
Report Date
October 18, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A REPORT OF A PATIENT WHO EXPERIENCED A SYSTEM ERROR 2240 ALARM THAT WAS CAUSED BY AN IMPROPER DISCONNECT/RECONNECT BY THE PATIENT DURING THERAPY. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY DISCONNECTING DURING EMERGENCIES. THE GUIDE ALSO PROVIDES INSTRUCTIONS FOR RETURNING TO THERAPY AFTER THE EMERGENCY DISCONNECT PROCEDURE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON THE HOMECHOICE DEVICE DURING DRAIN THREE OF FOUR OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, IT WAS NOTED THAT THE PATIENT HAD INCORRECTLY DISCONNECTED FROM THE SET AND THEN RECONNECTED THEMSELVES TO THE PATIENT LINE AFTER RECEIVING THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE CARE GIVER WITH CLEARING THE ALARM. PROPER PROCEDURES WERE REVIEWED AND THE CARE GIVER WAS ADVISED TO CONTACT THEIR NURSE FOR FURTHER MEDICAL DIRECTION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736904 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE