FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 4250210
·
Received November 6, 2014
Report
- Report Number
- 3008642652-2014-03703
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 11, 2014
- Report Date
- November 5, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (THERAPY ELECTRODE NON-FUNCTIONAL) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE BETWEEN THE FRONT TE AND ECG A. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE.
Description of Event or Problem · 1
A ZOLL DISTRIBUTOR RETURNED ELECTRODE BELT SN (B)(4) TO REPORT THAT THE THERAPY ELECTRODES WERE NON-FUNCTIONAL.