FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4250210 · Received November 6, 2014

Report

Report Number
3008642652-2014-03703
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 11, 2014
Report Date
November 5, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (THERAPY ELECTRODE NON-FUNCTIONAL) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE BETWEEN THE FRONT TE AND ECG A. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE.

Description of Event or Problem · 1

A ZOLL DISTRIBUTOR RETURNED ELECTRODE BELT SN (B)(4) TO REPORT THAT THE THERAPY ELECTRODES WERE NON-FUNCTIONAL.