FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4250171 · Received November 6, 2014

Report

Report Number
3008642652-2014-03680
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 9, 2014
Report Date
November 4, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) HAS BEEN CONFIRMED. THE CAUSE OF THE PROBLEM WAS ISOLATED TO FLASH MEMORY COMPONENTS U102 AND U105 ON COMPUTER ANALOG (CA) BOARD CONNECTION. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN ((B)(4)) WAS APPROVED BY FDA ON 4/16/2013. IMPLEMENTATION BEGAN ON 5/9/2013. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A ZOLL DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) TO REPORT THAT THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER ON.