FDA Adverse Event
Injury
Summary report: N
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
MDR report key: 4250049
·
Received November 6, 2014
Report
- Report Number
- 9710014-2014-00582
- Event Type
- Injury
- Date Received
- November 6, 2014
- Date of Event
- April 7, 2014
- Report Date
- October 30, 2014
- Manufacturer
- MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH
- Product Code
- MCM
- PMA / PMN Number
- P000025
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BASED ON THE RECEIVED INFORMATION, THE OBSERVED SYMPTOMS ARE VERY LIKELY DUE TO AN ELECTRODE MIGRATION OUT OF COCHLEA. THE PATIENT HAS A MONDINI MALFORMATION AND A FULL INSERTION COULD NOT BE ACHIEVED AT IMPLANTATION (TWO ELECTRODE CHANNELS WERE OUT OF COCHLEA). NOW SIX CHANNELS ARE REPORTEDLY OUT OF COCHLEA. A REPOSITIONING SURGERY IS TO BE DISCUSSED WITH THE SURGEON, BUT A DATE FOR REVISION SURGERY HAS NOT BEEN RECEIVED.
Description of Event or Problem · 1
THE PATIENT CAN ONLY DETECT LOUD SOUNDS. SPEECH SOUNDS ARE NOT RELIABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 716015 | MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM | SONATA | MCM | MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |