FDA Adverse Event Malfunction Summary report: N

COSYCOT INFANT WARMER

MDR report key: 4249356 · Received November 14, 2014

Report

Report Number
9611451-2014-00867
Event Type
Malfunction
Date Received
November 14, 2014
Report Date
October 16, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE HEAD WARMER ASSEMBLY OF THE COMPLAINT IW934 COSYCOT INFANT WARMER WAS RETURNED TO OUR SERVICE CENTER IN (B)(4), WHERE IT WAS INSPECTED BY A TRAINED FISHER & PAYKEL HEALTHCARE TECHNICIAN. RESULTS: VISUAL INSPECTION REVEALED A CRACK ON THE RETURNED HEAD WARMER ASSEMBLY. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 070712. CONCLUSION: IT IS LIKELY THAT THE PHYSICAL DAMAGE OF THE COMPLAINT WARMER HEAD IS RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE HEATER HEAD BEGINS TO WARM UP DURING USE A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRAZING AND CRACKING OF THE HEATER HEAD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNIT FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS THE POLYCARBONATE & ACRYLIC COMPONENTS AND STATES THE FOLLOWING: - "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE" - "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES." THE HEAD CASE AND HARNESS KIT WERE REPLACED ON THE COMPLAINT HEAD WARMER ASSEMBLY AND IT WAS RETURNED TO THE CUSTOMER AFTER PASSING THE SAFETY AND PERFORMANCE CHECKS.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6), REPORTED THAT THE HEATER HEAD CASE OF AN IW934 COSYCOT INFANT WARMER WAS CRACKED. A SERVICE OF THE DEVICE WAS REQUESTED. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736487 COSYCOT INFANT WARMER FMT FMT FISHER & PAYKEL HEALTHCARE LTD IW934 070712

Patients

Seq Age Sex Outcome Treatment
1