530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-50366
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- September 19, 2014
- Report Date
- October 14, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH CRACKED LCD WINDOW, CRACKED CASE AT LCD WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, SCRATCHES ON RESERVOIR TUBE WINDOW AND CRACKED BATTERY TUBE THREADS.
NO BLOOD GLUCOSE LEVELS REPORTED. THE CUSTOMER REPORTED A MOTOR ERROR ALARM FROM THE INSULIN PUMP DURING BASAL. TROUBLESHOOTING WAS DONE. IT WAS REPORTED THAT THE CUSTOMER WAS USING THE SENSOR FEATURE OF THE INSULIN PUMP. THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NOT EXPOSED TO ANY STRONG MAGNETIC FIELD. THE CUSTOMER WAS ABLE TO COMPLETE THE REWIND SEQUENCE ON THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 726231 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-551NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |