PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-50719
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
THE INSULIN PUMP PASSED THE FUNCTIONAL TESTS. HOWEVER, THE DEVICE HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO THE KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, AND A CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE CUSTOMER'S FATHER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR, NO DELIVERY AND CUSTOMER HAD HIGH BLOOD GLUCOSE. CUSTOMER STATED SHE GOT THE BUTTON ERROR WHEN SHE WAS GIVING BOLUS TO HERSELF. CUSTOMER STATED HAD HIGH BLOOD GLUCOSE AT THAT TIME BUT CANNOT RECALL THE BLOOD GLUCOSE READING. CUSTOMER TREATED WITH MANUAL INJECTION. CUSTOMER STATED THAT IN THE PAST HE WENT TO PRIME THE INSULIN PUMP AND GOT NO DELIVERY ALARM. CUSTOMER'S BLOOD GLUCOSE AT THAT TIME WAS AROUND 130 MG/DL. FATHER TOO THE SET OUT AND DID A REWIND AND WAS ABLE TO GET IT TO WORK. UNABLE TO TROUBLESHOOT DUE TO BUTTON ERROR. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725483 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |