HEARTWARE VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-01155
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- August 28, 2014
- Report Date
- September 30, 2014
- Manufacturer
- HEARTWARE
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE HVAD REMAINS IMPLANTED IN THE PATIENT, THEREFORE IT WILL NOT BE RETURNED. HOWEVER, THE BATTERIES ((B)(4)) ARE AVAILABLE FOR EVALUATION, BUT HAVE NOT BEEN RECEIVED BY THE MANUFACTURER. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION.
THE HVAD IS USED FOR TREATMENT NOT DIAGNOSIS. IT WAS REPORTED THAT PATIENT STATED THAT BATTERY ((B)(4)) DISPLAYED FULLY CHARGED WHEN IN HIS HAND, BUT WHEN CONNECTED TO THE CONTROLLER IT READS THAT IT ONLY HAS TWO (2) BARS. THE BATTERY ((B)(4)) STAYS CONNECTED FOR ROUGHLY FOUR (4) MINUTES THEN SWITCHES TO THE OTHER BATTERY. THIS WAS REPRODUCED BY THE PERFUSIONIST AND THE BATTERIES WERE REPLACED. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THIS EVENT. THE TWO BATTERIES WERE RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. ANALYSIS OF THE DEVICES REVEALED THAT THE BATTERIES MET SPECIFICATIONS; THE BATTERIES PASSED VISUAL INSPECTION AND FUNCTIONAL TESTING. THE REPORTED EVENT COULD NOT BE CONFIRMED VIA REVIEW OF THE CONTROLLER LOG FILES, BECAUSE THE AVAILABLE LOG FILES DID NOT COVER THE REPORTED EVENT. BASED ON THE RESULTS OF THE PREVIOUS MANUFACTURER'S INTERNAL INVESTIGATION, FIELD ACTIONS AND CHANGES TO THE BATTERY INTERNAL CELL SUPPLIER, NO ADDITIONAL INVESTIGATION OF THIS EVENT IS REQUIRED AT THIS TIME. THE MOST LIKELY CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO FAULTY INTERNAL CELLS WITHIN THE BATTERY PACK. Z-1607-2014. FIELD SAFETY NOTICE (FSCA APR2014) WAS ISSUED TO PROVIDE PATIENTS AND HEALTHCARE PROVIDERS WITH INFORMATION TO RECOGNIZE BATTERIES WITH LESS THAN TWO HOURS OF RUN TIME, AS WELL AS REEMPHASIZE INSTRUCTION ON ACTIONS TO TAKE WHEN BATTERY ALARMS OCCUR AND REINFORCE PROPER POWER MANAGEMENT. FIELD SAFETY NOTICE WAS THEN EXPANDED (FSCA APR2014.1) IN ORDER TO REMOVE BATTERIES FROM THE FIELD THAT WERE RELEASED PRIOR TO THE IMPLEMENTATION OF ENHANCED BATTERY SCREENING PROCESS TO ADDRESS AND PREVENT BATTERY FAILURES. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS, INCLUDING TRANSPORTATION. THE INSTRUCTIONS FOR USE (IFU), PATIENT MANUAL, AND TRAINING MATERIAL PROVIDE CLEAR INSTRUCTIONS TO THE USER ON PROPER USAGE AND CARE OF POWER SOURCES. THE IFU PROVIDES INSTRUCTION TO FURTHER EDUCATE THE PATIENT ABOUT PRODUCT SAFETY, VISUAL AND AUDIBLE ALARM MANAGEMENT, AND DEVICE MANAGEMENT; ADDITIONAL GUIDELINES INSTRUCT THE USER ON HOW TO DETECT AND REACT TO A POWER SOURCE MALFUNCTION. ADDITIONALLY THERE IS A WARNING TO KEEP SPARE, FULLY CHARGED BATTERIES AND BACK UP CONTROLLER AVAILABLE AT ALL TIME. THE STEPS FOR EXCHANGE OF BATTERIES AND CONTROLLERS ARE OUTLINED. HEARTWARE IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER FLA-14-14, DATED JUNE 2, 2014, AND PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. BATTERIES - (B)(4)/ 1650 EXPIRATION DATE: 31 OCT 2014 (BATTERIES RECEIVED (B)(4) 2014). (B)(4).
APPROXIMATELY NINE MONTHS POST HVAD IMPLANTATION, IT WAS REPORTED THAT THE PATIENT EXPERIENCED ISSUES WITH THE BATTERIES THAT INCLUDED ABNORMAL SWITCHING BETWEEN POWER SOURCES AND RAPID BATTERY DEPLETION. BOTH THE BATTERIES WERE REPLACED AND RETURNED TO HEARTWARE FOR EVALUATION. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THIS EVENT. INVESTIGATION IS ONGOING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 727474 | HEARTWARE VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM | DSQ | HEARTWARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) - BATTERY |