ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-32203
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Report Date
- October 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 02/12/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/29/2015 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED OR CONFIRMED. THE KEYPAD COVER WAS UNDAMAGED. THE KEYPAD BUTTONS WERE NORMALLY RESPONSIVE. THERE WAS CONTAMINATION FOUND UNDER THE CONTRAST BUTTON. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ALSO UNRELATED, THE BATTERY COMPARTMENT WAS CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE OK KEYPAD BUTTON WAS UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 727772 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |