FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4243496 · Received November 11, 2014

Report

Report Number
2531779-2014-32194
Event Type
Malfunction
Date Received
November 11, 2014
Report Date
October 22, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/15/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE ALARM HISTORY INDICATED AN OCCLUSION ALARM AT 01:00 ON (B)(6) 2014; DELIVERIES WERE RESUMED WHEN THE PUMP WAS PRIMED AT 13:35 THE SAME DAY. A REVIEW OF THE TOTAL DAILY DOSE HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THE PUMP SUCCESSFULLY COMPLETED A REWIND, LOAD, AND PRIME SEQUENCE. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ISSUES OCCURRING. THE PUMP PASSED DELIVERY ACCURACY TESTING AND WAS FOUND TO BE DELIVERING WITHIN REQUIRED SPECIFICATIONS. THE COMPLAINT OF INACCURATE DELIVERY COULD NOT BE CONFIRMED OR DUPLICATED ON INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014 THE REPORTER CONTACTED ANIMAS ALLEGING A NON-SPECIFIC INACCURATE DELIVERY ISSUE WITH THE PUMP. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLE SHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
726630 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 61 YR