ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-32185
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Report Date
- November 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON 11/3/2014 WITH THE FOLLOWING FINDINGS: DISPLAY IS DIM/FADING/REDDISH, BATTERY COMPARTMENT IS CRACKED BELOW BUMPER PAD, INVESTIGATION COMPLETED USING THE RETURNED BATTERY CAP, BATTERY CAP IS DAMAGED/STRIPPED. PUMP CASE WAS REMOVED AND CORROSION WAS FOUND UNDERNEATH THE INTERNAL BATTERY. BLACK BOX CONFIRMED TIME AND DATE RESET TO DEFAULT AFTER BATTERY CHANGE ON 09/09/2014, PUMP WAS EXERCISE FOR 24 HOURS, THEN TOOK OUT THE BATTERY FOR 6 HOURS AND THE TIME AND DATE RESET COMPLAINT WAS DUPLICATED ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A DIM DISPLAY, DAMAGED BATTERY COMPARTMENT AND CAP. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON 11/3/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 726050 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |