FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4243350 · Received November 11, 2014

Report

Report Number
2531779-2014-32185
Event Type
Malfunction
Date Received
November 11, 2014
Report Date
November 3, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON 11/3/2014 WITH THE FOLLOWING FINDINGS: DISPLAY IS DIM/FADING/REDDISH, BATTERY COMPARTMENT IS CRACKED BELOW BUMPER PAD, INVESTIGATION COMPLETED USING THE RETURNED BATTERY CAP, BATTERY CAP IS DAMAGED/STRIPPED. PUMP CASE WAS REMOVED AND CORROSION WAS FOUND UNDERNEATH THE INTERNAL BATTERY. BLACK BOX CONFIRMED TIME AND DATE RESET TO DEFAULT AFTER BATTERY CHANGE ON 09/09/2014, PUMP WAS EXERCISE FOR 24 HOURS, THEN TOOK OUT THE BATTERY FOR 6 HOURS AND THE TIME AND DATE RESET COMPLAINT WAS DUPLICATED ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A DIM DISPLAY, DAMAGED BATTERY COMPARTMENT AND CAP. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON 11/3/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
726050 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 17 YR