FLOGARD
MDR report key: 4243307
·
Received November 11, 2014
Report
- Report Number
- 1416980-2014-39883
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- October 14, 2014
- Report Date
- October 14, 2014
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION REVEALED A DAMAGED LATCH ROLLER. THE CAUSE OF THE CONDITION WAS NOT DETERMINED. TO CORRECT THE CONDITION, THE LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED LATCH ROLLER. NO ADDITIONAL INFORMATION IS AVAILABLE.