FDA Adverse Event Malfunction Summary report: N

LIGACLIP CLIP APPLIER

MDR report key: 4243273 · Received November 11, 2014

Report

Report Number
3005075853-2014-07805
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 17, 2014
Report Date
October 20, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS OF THE ER320 DEVICE FOUND THAT IT WAS RECEIVED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE MISALIGNED. A BAG WITH ONE SCISSORED CLIP WAS RETURNED ALONG WITH THE INSTRUMENT. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. THE DEVICE WAS CYCLED AND ELEVEN SCISSORED CLIPS WERE FED AND FORMED DUE TO THE MISALIGNED CONDITION OF THE JAWS. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY, SCISSORING OCCURRED. THE DEVICE WAS USED ON THE BLOOD VESSEL. THE DOCTOR COMMENTED THAT THERE MIGHT HAVE BEEN TORQUING OR TWISTING OF THE DEVICE PRESENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
725408 LIGACLIP CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA L4ET3X

Patients

Seq Age Sex Outcome Treatment
1