LIGACLIP CLIP APPLIER
Report
- Report Number
- 3005075853-2014-07805
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- October 17, 2014
- Report Date
- October 20, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS OF THE ER320 DEVICE FOUND THAT IT WAS RECEIVED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE MISALIGNED. A BAG WITH ONE SCISSORED CLIP WAS RETURNED ALONG WITH THE INSTRUMENT. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. THE DEVICE WAS CYCLED AND ELEVEN SCISSORED CLIPS WERE FED AND FORMED DUE TO THE MISALIGNED CONDITION OF THE JAWS. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY, SCISSORING OCCURRED. THE DEVICE WAS USED ON THE BLOOD VESSEL. THE DOCTOR COMMENTED THAT THERE MIGHT HAVE BEEN TORQUING OR TWISTING OF THE DEVICE PRESENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725408 | LIGACLIP CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | L4ET3X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |