G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-25832
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- October 16, 2014
- Report Date
- October 16, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED AND FOUND NO DEFECTS. RECEIVER DATA LOG WAS DOWNLOADED AND NO ERRORS WERE OBSERVED. DEVICE FAILED FUNCTIONAL TESTING. FURTHER TESTS SHOWED THAT THE SPEAKER IS DEFECTIVE. THE ROOT CAUSE WAS DETERMINED TO BE AN ASSEMBLY ERROR. A CAPA HAS BEEN INITIATED FOR THIS ISSUE.
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO AUDIO OUTPUT PATIENT EXPERIENCED ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT'S MOTHER TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE NOT FUNCTIONAL, BUT DEVICE DID VIBRATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725840 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |