FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4242829 · Received November 11, 2014

Report

Report Number
2032227-2014-49919
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 12, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE REPORTER DID NOT KNOW THE BLOOD GLUCOSE READING. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
725187 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 38 YR