MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-05645
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- September 16, 2014
- Report Date
- October 14, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE INSTRUMENT PITCH CABLE WAS BROKEN AT THE PROXIMAL CLEVIS HUB. THE BROKEN STRANDS STUCK OUT AT THE INSTRUMENT WRIST. NO OTHER DAMAGE OR WEAR MARKS WERE OBSERVED ON THE CLEVIS. THE OTHER CABLES AT THE WRIST WERE INTACT AND UNDAMAGED. INTUITIVE SURGICAL, INC. (ISI) HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WOULD AFFECT ANY MATERIAL OF THE FINAL PRODUCT AND/OR THE QUALITY OR PERFORMANCE OF THE INSTRUMENT USED DURING THIS REPORTED EVENT. THE REPORTED COMPLAINT THAT THE INSTRUMENT WAS SPASTIC WITH STRANGE MOVEMENT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT. RECURRENCE OF THE ALLEGED FAILURE MODE IS NOT EXPECTED TO LIKELY CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY OR SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT. HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT BEHAVED LIKE A SPASTIC ARM, AND STRANGE MOVEMENT WAS REPORTED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725634 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-09 | N10140407 015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |