FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4242802 · Received November 11, 2014

Report

Report Number
2032227-2014-49745
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 9, 2014
Report Date
October 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. HOWEVER, THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AND ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A FAULTY FORCE SENSOR RESISTOR. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING PRIME. CUSTOMER STATED THAT THEY WERE AT A FAIR PRIOR TO THE ALARM. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND THEY ARE ABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 123 MG/DL. NO FURTHER INFORMATION REPORTED.