ACTIVE CORD
Report
- Report Number
- 3005099803-2014-03547
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Report Date
- October 20, 2014
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- KNS
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE COMPLAINANT WAS UNABLE TO PROVIDE THE LOT NUMBER. THEREFORE, THE MANUFACTURE DATE AND EXPIRATION DATE ARE UNKNOWN. REPORTED EVENT OF CORD BROKEN. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
VISUAL EVALUATION OF THE COMPLAINT DEVICE FOUND THE FEMALE CONNECTOR TO BE DETACHED FROM THE CORD. ADDITIONALLY, THE CORD APPEARS TO HAVE BEEN BENT ADJACENT TO THE EDGE OF THE CONNECTOR PRIOR TO FAILURE. THE COMPLAINT WAS CONFIRMED. THE FEMALE CONNECTOR WAS FOUND TO BE DETACHED WHICH IS CONSISTENT WITH THE COMPLAINT INDICATING THAT THE ACTIVE CORD WAS BROKEN. HOWEVER, THE MOST PROBABLE ROOT CAUSE OF THIS EVENT IS UNDETERMINABLE AS AN EXAMINATION OF THE DEVICE, EVENT DESCRIPTION, AND FOLLOW-UP INFORMATION DID NOT PROVIDE ENOUGH INFORMATION TO DETERMINE A ROOT CAUSE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ACTIVE CORD VALLEY LAB WAS USED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE ACTIVE CORD BROKE. DESPITE NUMEROUS ATTEMPTS BOSTON SCIENTIFIC HAS BEEN UNABLE OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ACTIVE CORD VALLEY LAB WAS USED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE ACTIVE CORD BROKE. DESPITE NUMEROUS ATTEMPTS BOSTON SCIENTIFIC HAS BEEN UNABLE OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725062 | ACTIVE CORD | UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | BOSTON SCIENTIFIC - SPENCER | M00561270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |