FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR HANDLE 33CM

MDR report key: 4242748 · Received November 11, 2014

Report

Report Number
0002936485-2014-00898
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 20, 2014
Report Date
October 20, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. THE FOLLOWING REPAIR AND SERVICE DIAGNOSTIC CODES WERE IDENTIFIED: COATING PEELING OFF HANDLE. REPLACED HANDLE. THE FOLLOWING WAS OBSERVED: COATING PEELING OFF THE HANDLE. THE HANDLE WAS REPLACED. THIS DOES CONFIRM THE ALLEGED FAILURE MODE OF INSULATION DAMAGE. PROBABLE ROOT CAUSE: POOR AUTOCLAVE RELIABILITY. INCORRECT STERILIZATION/REPROCESSING PROCEDURE. HANDLING PROCEDURES. CONTACT FORCES. PRODUCT USED BEYOND DEFINED USEFUL LIFE. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.