FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4242747 · Received November 11, 2014

Report

Report Number
2032227-2014-50211
Event Type
Injury
Date Received
November 11, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE BASIC OCCLUSION TEST AND THE DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A FAULTY FORCE SENSOR RESISTOR. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS, A BROKEN RESERVOIR TUBE LIP AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 461 MG/DL. THE CUSTOMER TREATED WITH A BOLUS. THE CUSTOMER WAS EXPERIENCING THE FOLLOWING SYMPTOMS: HEADACHE AND STRESS. THE CUSTOMER'S SENSOR READ 293 MG/DL AS HER BLOOD GLUCOSE LEVEL. HER BLOOD GLUCOSE LEVEL AT THE TIME OF THE REPORT WAS 123 MG/DL. THE CUSTOMER DECLINED TROUBLESHOOTING AND STATED SHE WOULD CHANGE HER INFUSION SET. THE CUSTOMER WAS ADVISED TO NOTIFY HER HEALTH CARE PROVIDER OF HER HIGH BLOOD GLUCOSE LEVEL. THE PRODUCT IS NOT BEING RETURNED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
725496 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 65 YR Other