G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-25831
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- September 30, 2014
- Report Date
- October 16, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. A TEST ALARM AND INTERIOR INSPECTION WAS PERFORMED AND THE DEVICE FAILED, CONFIRMING THE REPORTED EVENT OF AN INTERMITTENT AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE AN ASSEMBLY ERROR.
(B)(4).
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM INTERMITTENT AUDIO OUTPUT ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE FUNCTIONAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 725438 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |