FDA Adverse Event Malfunction Summary report: N

TPS MICRO DRIVER

MDR report key: 4242486 · Received November 11, 2014

Report

Report Number
0001811755-2014-04032
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 15, 2014
Report Date
October 15, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K943540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON EVALUATION A DAMAGED PCB WAS DISCOVERED, WHICH CAN CAUSE OR CONTRIBUTE TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED DURING A PROCEDURE, THE TPS MICRO DRIVER WAS RUNNING IN SAFE MODE. THE DEVICE ALSO WOULD NOT GO IN THE CORRECT INTENDED DIRECTION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. THE CASE WAS COMPLETED WITH THE SAME DEVICE WITHOUT ANY DELAY.

Description of Event or Problem · 1

IT WAS REPORTED DURING A PROCEDURE, THE TPS MICRO DRIVER WAS RUNNING IN SAFE MODE. THE DEVICE ALSO WOULD NOT GO IN THE CORRECT INTENDED DIRECTION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. THE CASE WAS COMPLETED WITH THE SAME DEVICE WITHOUT ANY DELAY.