FDA Adverse Event Malfunction Summary report: N

DUAL TRIGGER ROTARY

MDR report key: 4242372 · Received November 11, 2014

Report

Report Number
0001811755-2014-04018
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
October 9, 2014
Report Date
October 14, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, A FAILED E-BOX WAS FOUND, WHICH CAN LEAD TO THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
727390 DUAL TRIGGER ROTARY INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT HWE STRYKER INSTRUMENTS-KALAMAZOO