FDA Adverse Event
Malfunction
Summary report: N
DUAL TRIGGER ROTARY
MDR report key: 4242372
·
Received November 11, 2014
Report
- Report Number
- 0001811755-2014-04018
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 14, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED EVENT WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, A FAILED E-BOX WAS FOUND, WHICH CAN LEAD TO THE REPORTED EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 727390 | DUAL TRIGGER ROTARY | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |