LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT
Report
- Report Number
- 2955842-2014-05641
- Event Type
- Malfunction
- Date Received
- November 11, 2014
- Date of Event
- September 10, 2014
- Report Date
- October 16, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A BROKEN PITCH CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE CABLE ON THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT SNAPPED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 727208 | LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420296-04 | M10131219 807 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |