FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 4242302 · Received November 11, 2014

Report

Report Number
2955842-2014-05641
Event Type
Malfunction
Date Received
November 11, 2014
Date of Event
September 10, 2014
Report Date
October 16, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A BROKEN PITCH CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE CABLE ON THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT SNAPPED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
727208 LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-04 M10131219 807

Patients

Seq Age Sex Outcome Treatment
1