FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4242133 · Received November 10, 2014

Report

Report Number
2531779-2014-32163
Event Type
Malfunction
Date Received
November 10, 2014
Report Date
October 28, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. NO DAMAGE WAS FOUND TO THE KEYPAD COVER. THE CONTRAST KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE, WHICH HAS NO EFFECT ON INSULIN DELIVERY. THE KEYPAD COVER WAS REMOVED, AND CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. UNRELATED TO THESE ISSUES, THE COLOR ON THE CASING WAS FADED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM AND DISCOLORED DISPLAY AND CONTAMINATION UNDER KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/28/2014.