ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-32146
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Report Date
- October 16, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/24/2015 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS THINNING. ALL OF THE KEYPAD BUTTONS WERE UNRESPONSIVE AND DIFFICULT TO PRESS. CONTAMINATION WAS FOUND UNDERNEATH THE UP, DOWN,A ND CONTRAST KEYPAD CONTACTS. UNRELATED TO THE KEYPAD ISSUE, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, A REPORTER CONTACTED ANIMAS ALLEGING THAT THE UP BUTTON WAS UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722875 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |