ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-32107
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Report Date
- October 25, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
THE METER REMOTE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE METER REMOTE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE METER REMOTE WAS POWERED ON AND IMMEDIATELY EMITTED AND ERROR 1 WITH SUB CODE 7. THE METER REMOTE COULD NOT BE PAIRED WITH A TEST PUMP DUE TO THE ERROR. THE ERROR COULD NOT BE CLEARED, INDICATING CORRUPTED DATA.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A METER ERROR (ERROR 1) ISSUE. IT WAS REPORTED THAT THE METER REMOTE EMITTED AN ERROR 1 WITH SUB CODE 7 DURING AN ATTEMPTED BOLUS, AND THE ERROR ALLEGEDLY DID NOT CLEAR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722814 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR |