FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4241980
·
Received November 10, 2014
Report
- Report Number
- 1416980-2014-39797
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 17, 2014
- Report Date
- October 17, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. THE BROKEN DOOR WAS IDENTIFIED DURING VISUAL INSPECTION. THE CAUSE OF THE CONDITION WAS UNKNOWN. TO CORRECT THE CONDITION, THE PUMP HEAD DOOR ASSEMBLY AND DOOR LATCH WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A BROKEN DOOR. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.