FDA Adverse Event Malfunction Summary report: N

COLIBRI

MDR report key: 4241953 · Received November 10, 2014

Report

Report Number
8030965-2014-10624
Event Type
Malfunction
Date Received
November 10, 2014
Report Date
October 29, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE CONTROL UNIT WAS NOT FUNCTIONING. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE SMALL BATTERY DRIVE DEVICE HAD ISSUES WITH THE CONTROLS. THE REPORTER INDICATED THAT THE BACKWARDS SPIN OPTION DID NOT FUNCTION AND THE SWITCH/BUTTON WAS JAMMED. ACCORDING TO THE REPORTER, WHEN THE SWITCH WAS PUSHED BACKWARDS, THE SPIN RAN, HOWEVER, THE DEVICE COULD NOT BE SWITCHED OFF OR STOPPED. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723298 COLIBRI INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF NA

Patients

Seq Age Sex Outcome Treatment
1