PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-50182
- Event Type
- Malfunction
- Date Received
- November 10, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, BROKEN BATTERY TUBE THREADS, CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE WINDOW, AND STAINED END CAP STICKER.
(B)(4) . CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED HE CHANGED THE BATTERY ON THE INSULIN PUMP, A CORNER JUST FEEL OFF THE SIDE. CUSTOMER'S BLOOD GLUCOSE WAS 160 MG/DL. THE BATTERY COMPARTMENT PIECE WAS MISSING WAS ABOUT 4 CM AND HAS TAPE CURRENTLY TO HOLD IT. CUSTOMER DID NOT WANT TO OPEN BATTERY COMPARTMENT. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 724238 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |