FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4241862 · Received November 10, 2014

Report

Report Number
2032227-2014-50107
Event Type
Malfunction
Date Received
November 10, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED UNEXPLAINED HIGH AND LOW BLOOD GLUCOSE. HE STATED HIS BLOOD GLUCOSE WILL BE HIGH WHEN HE WAKES UP, SOMETIMES IN THE 500S. HE DOES NOT KNOW WHY IT IS HIGH. CUSTOMER DECLINED SPEAKING WITH HELP LINE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723564 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR Required Intervention